Caple C196/54 Specifikace Strana 35

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 36
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 34
35
High Performance Liquid Chromatograph
P. 14
Light source
Number of diode elements
Wavelength range
Bandwidth, slit width
Wavelength accuracy
Wavelength precision
Noise
Drift
Linearity
Functions
Cell
Cell temperature control range
Web control
Buffer memory
Operating temperature range
Dimensions, weight
Power requirements
UV-VIS Detectors SPD-20A / 20AV
Photodiode Array Detector SPD-M20A
SPD-M20A (228-45005-XX)
512
190 to 800 nm
1.2 nm (high-resolution mode),
8 nm (high-sensitivity mode)
0.6 × 10
−5
AU (under specified conditions)
5 × 10
−4
AU/h (under specified conditions)
2.0 AU (ASTM standard)
Contour output, spectrum library,
MAX plotting
Optical wavelength: 10 mm,
Capacity: 10 µL, Pressure: 12 MPa
Parameter setting, log management,
management of consumable parts, etc.
Approx. 20 minutes of data in
the entire wavelength region
(only when using LabSolutions)
W260 × D420 × H140 mm, 12 kg
AC 110 V, 230 V, 150 VA, 50/60 Hz
SPD-20AV (228-45004-XX)
190 to 900 nm
± 1 nm max.
± 0.1 nm max.
5°C above room temperature to 50°C
4 to 35°C
SPD-20A (228-45003-XX)
Deuterium (D
2
) lamp
190 to 700 nm
Deuterium (D
2
) lamp, tungsten (W) lamp
None
8 nm
0.5 × 10
−5
AU (under specified conditions)
1 × 10
−4
AU/h (under specified conditions)
2.5 AU (ASTM standard)
Dual-wavelength detection in the range 190 to 370 nm and upwards of
371 nm, ratio-chromatogram output, wavelength scanning
Optical wavelength: 10 mm,
Capacity: 12 µL, Pressure: 12 MPa
Refer to the information on the CBM-20A/20Alite
W260 × D420 × H140 mm, 13 kg
AC 110 V, 230 V, 160 VA, 50/60 Hz
Cell
Op
SPD-20A
SPD-M20A
* Prominence UFLC and Prominence UFLC
XR system use SPD-20A UFLC version (228-45130-xx) which has a semi-micro temperature-controlled flow cell installed
for optimization of fast analysis. (Standard type temperature-controlled flow cell is optional for SPD-20A UFLC version.)
Light source
Wavelength range
Spectral bandwidth
Wavelength accuracy
Wavelength reproducibility
S/N
Cell
(capacity, pressure resistance, material)
Cell Temperature input range
Cell
Functions
Safety measures
Operational temperature range
Dimension, weight
Power requirements
RF-20A (228-45147-XX)
Xenon lamp
0, 200 to 650 nm
Water Raman peak S/N 1200 min.
Any two wavelengths between 200 and 650 nm
W260 × D420 × H210 mm, 16 kg
RF-20A
XS (228-45148-XX)
Xenon lamp, low-pressure mercury lamp
(to check wavelength accuracy)
0, 200 to 750 nm
Water Raman peak S/N 2000 min.
4 to 40°C, 1°C step
(Room temperature − 10°C) to 40°C
(2 mL/minute max. flow rate, 85°C max. oven temperature)
Any two wavelengths between 200 and 750 nm
W260 × D420 × H210 mm, 18 kg
20 nm
± 2 nm
± 0.2 nm
12 µL; 2 MPa (approx. 20 kgf/cm
2
); SUS316L, PTFE (fluororesin), quartz
0.5 s per wavelength
4 to 35°C
AC 110 V, 230 V, 400 VA, 50/60 Hz
P. 16
W
C
ell
(cap
Cel
Ope
Fluorescence Detectors RF-20A / 20AXS
P. 18
RID-20A (228-45104-XX)
1 to 1.75 RIU
2.5 × 10
−9
RIU max.
1 × 10
−7
RIU/h max.
A mode: 0.01 × 10
−6
to 500 × 10
−6
RIU, P and L modes: 1 × 10
−6
to 5,000 × 10
−6
RIU
0.05 to 10 sec, 10 steps
Supported
Auto zero, optical zero, fine zero
20 mL/min (150 mL/min with option)
30 to 60°C
9 µL
2 MPa (cell unit)
4 to 35°C
W260 × D420 × H140 mm, 12 kg
AC 110 V, 230 V, 150 VA, 50/60 Hz
Refractive index range
Noise level
Drift
Range
Response
Polarity switching
Zero adjustment
Maximum operating flow rate
Temperature control of cell unit
Cell volume
Cell withstand pressure
Operating temperature range
Dimensions, weight
Power requirements
Refractive Index Detector RID-20A
Zobrazit stránku 34